Specifications; Maintenance; Cleaning And Storage - Extech Instruments CO210 User Manual

Co2 monitor and datalogger
Hide thumbs Also See for CO210:
Table of Contents

Advertisement

Specifications

Function
CO
2
Temperature
Humidity
Display
Sensor Type
Response
Datalogger
Operating Conditions
Storage Conditions
Power Supply
Dimensions
Weight

Maintenance

Cleaning and storage

1. The meter should be wiped with a damp cloth and mild detergent when necessary. Do not
use solvents or abrasives.
2. Store the meter in an area with moderate temperature and humidity.
Range
0 to 9999ppm
-10 to 60°C (14 to 140°F)
0.1 to 99.9%
Multi-function LCD
CO
Humidity: Capacitance sensor
Temperature (air): Thermistor
CO
Air temperature: < 2min for 90% step change
RH: < 10min for 90% step change
15999 total samples (5,333 readings each for CO
temperature, and relative humidity).
To download logged readings, use the latest Windows-
based PC software and USB driver from the Extech
website. USB cable is included with the CO210.
14 to 140°F (-10 to 60°C); < 90% RH non-condensing
-4 to 140°F (-20 to 60°C); <99% RH non-condensing
5VDC (±10%), ≥ 500mA (supplied adaptor)
4.6 x 4 x 4" (117 x 102 x 102 mm)
7.2 oz (204 g)
Resolution
1ppm
0.1°
0.1%
: NDIR (non-dispersive infrared) technology
2
: < 2min for 90% step change
2
8
Accuracy
± (5%rdg + 50ppm)
± 0.6°C (0.9°F)
± 3% (10 to 90%)
± 5% (< 10% or > 90%)
, air
2
CO210-en-US v1.3 1/21

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CO210 and is the answer not in the manual?

Questions and answers

Table of Contents