Final inspection

Check to ensure that no parts have been left in the packaging.
Inspect the lift for signs of wear and damage.
Check all four castor wheels and castor wheel locks.
Check all connections and fixtures including screws and bolts.
Eva450/Eva600
Eva400
Check the emergency stop function by activating the emergency stop button, and
then pressing either the up or down button. If nothing happens when the up or
down buttons are pressed, the emergency stop is functioning properly.
Grasp the hand control, press the up button and run the lift arm all the way up.
Then, press the down button and run the lift arm all the way down.
Test base-width adjustment function. Press the button for base-width adjustment
to widen the base fully, and then press the other button to narrow the base again.
EM
Test lift function by lifting a person (not a user) using an approved sling. At the
EE
same time, check the emergency lowering function with someone on the lift. See
section on Emergency lowering.
If the lift is functioning correctly, connect the charger and check to ensure that the
charging lamp on the control box lights up.
NOTE!
Before the lift is used for the first time:
• it must be charged for at least 4 hours. See section on charging batteries.
• the control box service counter needs to be reset. To reset the service counters
Eva450/Eva600
press both lift buttons on the hand control at the same time for 5 seconds.
An audio signal will indicate that the timer has been reset.
Keep the user manual where it is accessible to users of the product.
SystemRoMedic
3
U S E R M A N U A L