Final Inspection - Handicare SystemRoMedic MiniLift160EE Manual

Table of Contents

Advertisement

Final inspection

Inspect the lift for signs of wear and damage. Check all castors and castor brakes.
Check all connections and fixtures including screws and bolts. Check the packaging to
ensure that there are no loose parts.
Check the emergency stop function by activating the emergency stop button, and
then pressing either the up or down button. If nothing happens when the up or down
buttons are pressed, the emergency stop is functioning properly.
Grasp the hand control, press the up button and run the lift arm all the way up. Then,
press the down button and run the liftarm all the way down.
Test base-width adjustment function by pushing down on the respective pedals for
widening and narrowing the base. For EE use the buttons on the hand control.
EE
Test lift function by lifting a person (not a user) using an approved sling. At the same
time, check the emergency lowering function with someone in the lift, see section on
Emergency lowering.
EM
If the lift is functioning correctly, connect the charger and check to ensure that the
charging lamp on the control box lights up.
NOTE!
Before the lift is used for the first time:
• it must be charged for at least 4 hours. See section on charging batteries.
• the control box service counter needs to be reset. To reset the service counters
press both lift buttons on the hand control at the same time for 5 seconds.
An audio signal will indicate that the timer has been reset.
Keep the user manual where it is accessible to users of the product.
SystemRoMedic
5
M A N U A L
TM

Advertisement

Table of Contents
loading

Table of Contents