Changing spare parts

80100001, -02 Front/rear castor 100 mm
Using an Allen key, remove the screw that holds the castor in place. Replace the castor, apply Loctite 2701 or similar,
and then secure the castor with the screw.
80100126 Front castor EML
Using an Allen key, remove the screw that holds the castor in place. Replace the castor, apply Loctite 2701 or similar,
and then secure the castor with the screw.
80100055, -56 Actuator leg legspreading
Connect the motor cable to the actuator.
Run the actuator until the legs are drawn in fully. Activate emergency stop. Disconnect the actuator from the control box.
Loosen and remove the screw that holds the actuator to the leg. Remove the screw covers and loosen the screw and
nut. Remove the actuator.
Install the new actuator; start by securing the end that is secured to the leg. Use the accompanying lock nut, and then
secure the end that is connected to the base of the mobile lift. Use the accompanying lock nut. Replace the screw
covers. Connect the actuator to the control box.
80100045, -46, -47, -48 Charger
Connect the charger cable to the control box, and then plug it into a wall socket.
80100032 Lever
Unscrew the old lever and install the new one.
80100044, -062, -063, -156 Hand control
Activate the emergency stop, and then disconnect the hand control. Connect the new hand control.
80100130 Hand control tester
80100053, -054 Actuator mast
Lower the mast to the lowest position. Activate emergency stop. Disconnect the actuator from the control box. Free the
upper end of the actuator by first removing the screw covers, and then removing the screws and nuts. Then, free the
lower end in the same way. Lift the mast and remove the actuator. Install the new actuator. Start by fitting the lower end,
and then lift the mast and fit the upper end. Secure the actuator with the existing screws and the new, accompanying
lock nuts. Replace the screw covers and connect the actuator to the control box.
80100058 Slingbar lock for slingbar alu
Remove the screws on either side of the safety latch and remove it. Install the new safety latch and secure it with the two
accompanying M5x12 screws. Position the safety latch, so that the bevelled side is facing in towards the lift hook.