ResMed Nox T3s How To Use Manual page 45

Noxturnal version 6
Table of Contents

Advertisement

Use the recording device log
After opening a recording with Noxturnal, click on the patient icon
The following chart will help you understand and identify the different lines in this log.
15:50
Oximeter Device On, event Oximeter connected. ID = 086BD71FADF4.
Oximeter Unstable, event Oximeter bonding success.
15:50
ID = 086BD71FADF4.
15:50
Log, eventOximeter configured.
15:50
Display Turned Off, event Display turned off
Recording Started, event Recording started. Data file: 01_072DB.NIF.
23:00
Last auto-zero calibration: Tue Jun 2 15:48:37 2020
23:00
Oximeter Device On, event Oximeter connected. ID = 086BD71FADF4.
Oximeter Unstable, event Oximeter bonding success.
23:00
ID = 086BD71FADF4.
23:00
Log, event Oximeter configured.
07:00
Recording Stopped, event Recording stopped. Data file: 01_072DB.NIF.
Abnormal lines or warnings
23:19
Log, event Oximeter bonding fail. ID = 086BD71FD4C2.
Oximeter Device Off, eventOximeter disconnected.
03:01
ID = 086BD71FD4C2.
06:13
Oximeter Battery Low, event Oximeter battery low.
00:45
Battery Status, event Battery status = 30%.
03:57
Low Battery, event Battery is empty! 1.
Message
and then on
Recording device.
Oximeter is connected, pairing process was successful
Nox T3s screen turnsoff
Recording starts
Oximeter is connected, pairing process was successful
Recording stops
Oximeter failed to connect
Oximeter turned off and disconnected
Oximeter batteries are almost empty
The Nox T3s battery is almostempty (notification at 30,
20, 10 and 0%)
TheNox T3s battery went flat before the end of the
recording
Meaning
Meaning