Cleaning and disinfection
Disinfecting the Nox T3s and accessories
Materials and equipment needed:
• Lint-free cloths
• Gloves
• PDI Sani-Cloth Plus Germicidal Disposable Cloth or equivalent validated disinfectant
Disinfection procedure:
1.
For disinfection, take a new wipe of PDI Sani-Cloth Plus Germicidal Disposable Cloth or equivalent validated disinfectant
2.
Wipe all surface areas of the component with the disinfectant for at least three minutes. If other disinfection materials are used
than PDI Sani-Cloth Plus Germicidal Disposable Cloth make sure:
• they are safe to use on metals and plastics
• to read the instructions from the manufacturer regarding required contact time of the solution to provide sufficient disinfection
3.
Allow components to air dry completely before next use (minimum 1 minute)
4.
Visually inspect the components under adequate lighting conditions (and magnification if needed) to confirm that the cleaning/
disinfection process has not damaged components. Inspect for surface wear, discoloration, corrosion, or cracking.
If necessary, the clips on the Nox T3s recorder can be removed before cleaning. The cleaning process described above also applies for
the clips. If the clips are visibly contaminated, they should be replaced.
* Equivalent validated disinfectant may be used if they are safe to use on gold plating, metals, and plastics. Similar disinfection wipes from PDI that are recommended to use with the Nox T3s System: Super Sani-Cloth Plus disinfection
wipes and Sani-Cloth AF Universal - Alcohol free Disinfection wipes (from PDI).
** If any component damage occurs during cleaning process, contact your local representative. Do not attempt to use the Nox T3s System until the device has been inspected and repaired by authorized Nox Medical personnel.
*
*
**