Patient hook-up
• Therapy device option: Braebon differential pressure sensor installation
Make sure you use an appropriate recording type.
Equipment needed:
- Differential pressure sensor kit (PN PN 7078820)
- 2 hydrophobic filters for each recording (PN 7073746)
- 2 AAA batteries for the differential pressure sensor (last for 30 to 50 recordings of 8 hours)
Install the differential pressure sensor
Make sure that 2 fresh AAA batteries are inserted and that you are using 2 new filters.
1 - Connect the green part of the pneumoflow sensor to the therapy device output.
2 - Connect the transparent part of the pneumoflow sensor to the patient tubing.
3 - Connect the red cable to input 1 or 2 of the Nox T3s.
4 - Connect the white cable to the black and red cable and plug it into the Airflow input
of the differential pressure sensor.
5 - Put the gain of the differential pressure sensor on LOW.
6 - Switch the differential pressure sensor ON just before the recording. It will automatically stop after
10 hours of recording.
Set-up picture:
20