Oximeter pairing fails during configuration
If the oximeter pairing process fails during the configuration, you can follow the instructions below:
1.
Check the BDA number in the Configuration window and make sure it matches the number on the oximeter.
2.
Restart the oximeter by removing the batteries and inserting batteries again to the oximeter. Then press the grey activation button
to turn on the device.Resume the pairing process.
3.
Restart the Nox T3s by unplugging the Nox T3s from the computer and plugging it back in again. Resume the pairing process.
43