Cleaning and disinfection
Cleaning the Nox T3s and accessories
All reusable components should be cleaned immediately after use to prevent accumulation of residual soil and minimize soil transfer
between patients. The Nox disposable RIP belts, nasal cannulas and filter tube connectors are single patient use only.
Materials and equipment needed:
• Endozime
AW Plus
®
• Lint-free cloths
• Gloves
• Soft bristle nylon brush (i.e. electrode brush, toothbrush, or nail brush)
Cleaning procedure:
1.
Prepare a solution with the Endozime
cleaner
2.
Dampen a lint-free cloth with the solution
• Do not pour or spray any liquids on the Nox T3s Recorder
• Do not allow any liquids to enter any openings on the Nox T3s Recorder
• Do not immerse the cables in liquid
• Avoid contact of the cleaning solution to cable/electrode connectors
3.
Wipe all surface areas thoroughly to remove all visible soil and contaminants. Wipe the component for at least 2 minutes. Use a
soft bristle nylon brush if necessary.
4.
Visually inspect the cleaned components to ensure no soil remains. Pay good attention to all junctions and details.
Repeat steps 2 and 3 if necessary.
5.
Allow components to air dry completely before disinfection (minimum of 3 minutes)
AW Plus hospital grade cleaner. Follow the instructions accompanying the hospital grade
®