Operation Example - Keysight Technologies N1913PM5B Product Note

Calorimetric power meter
Table of Contents

Advertisement

Operation Example

1. Refer to the VDI PM5B Operation Manual for details. See page 3 for a link.
2. Load the software (optional - software not required for manual operation).
3. On the front panel, set the Range Switch to Remote and Cal Factor Switch
to 00.0 (Figure 3).
Figure 3
PM5B Power Meter Front Panel
Display
Reads power on
selected range.
Sensor Head
Connector Port
For use with
standard Amphenol
cable.
4. Turn on the PM5B power meter using the power ON/OFF switch on the rear panel
(Figure 4).
5. Start the PM5B control program (optional - not required with manual operation).
6. Enter serial number, for example "520V" (optional - not required with manual operation).
Figure 4
PM5B Power Meter Rear Panel
Internal
Calibration
Switch
(100 µW to
100 mW)
Analog Output
BNC to other
equipment.
(-10 to +10 VDC)
Keysight N1913PM5B Product Note
Zero Button
Press to re-zero display
Calibration Input
Banana jacks for DC calibration. (max 15V)
DC Power Input
(9V, 4A)
Range Switch
Select ranges from
200 µW to 200mW.
'Remote' enables
USB control and
auto-ranging.
Calibration Factor
Switch (dB)
Used to adjust the
power meter display.
Power ON/OFF
Switch
Toggle switch.
USB Interface
Type B jack.
7