CAIRE FreeStyle Manual page 48

Portable oxygen concentrators with ultrasense and patient-removable
Hide thumbs Also See for FreeStyle:
Table of Contents

Advertisement

1. Turn unit off and disconnect the power cord.
2. Remove the back housing, and locate the product filter/volume tank. See figure below.
3. Disconnect tubing from the inlet of the product filter/volume tank.
4. Unscrew the product filter/volume tank from the product manifold
5. Place the o-ring into the o-ring groove located on top of the volume tank.
6. Fasten the new product filter/volume tank to the product manifold.
7. Connect the tubing and secure with a tie-wrap.
If the compressor inlet felt filter (shown above) is contaminated, a
replacement felt filter can be ordered by contacting CAIRE and
referencing Part Number
INFECTION CONTROL
With growing concern about possible cross infection from home oxygen equipment from one
home care patient to another, a clarification on this topic is necessary. The organisms of most
concern are M. Tuberculosis, HIV, and Viral Hepatitis. These are potentially pathogenic.
Portable Oxygen Concentrator
Figure 9: Product Filter
FI143-2
Product
Filter
Compressor
Felt Filter
MN220-1 B 48