Wldwin Error Message - SubSurface Instruments LD-18 Operation Manual

Digital noise reduction water leak detector
Table of Contents

Advertisement

5. WLDwin error message

Message
"Connect the detector!"
Meaning
Connected detector cannot be confirmed at the time of execution of detector edit menu.
Cause and
Check connection between PC and detector, and check if the power for the detector is ON.
remedy
* See Software Volume "3-2 Check of communication with detector" (page39).
Check if the dedicated driver has been correctly installed.
* See Software Volume "3 Installation of driver" (page36).
Message
"No response from detector. Interrupted!"
Meaning
Communication was interrupted due to any reason during communication with the detector.
Cause and
Check connection between PC and detector, and check if the power for the detector is ON.
remedy
* See Software Volume "3-2 Check of communication with detector" (page39).
Check the battery capacity of the detector.
* See Hardware Volume "3-2 Check of battery capacity" (page12).
Message
"Data read ended in failure!"
Meaning
Failure in data read
Cause and
Check connection between PC and detector, turn ON the power for the detector again,
remedy
and execute data read.
Message
"Data clear ended in failure!"
Meaning
Failure in data read
Cause and
Check connection between PC and detector, turn ON the power for the detector again,
remedy
and execute data clear.
Message
"Setup ended in failure!"
Meaning
Failure in setup of detector date / hour
Cause and
Check connection between PC and detector, turn ON the power for the detector again,
remedy
and execute date / hour setup.
Message
"Data is incorrect!
There is a possibility where displayed data is erroneous."
Failed in read of saved file.
Meaning
Cause and
There is a possibility where the saved file was broken.
remedy
If data in the saved file was broken, re-edit with this software is not permitted.
Message
"Maximum number of data was exceeded!
The number of data is 250 at maximum."
Number of graph display data is excessive at the time of data add or data move.
Meaning
Cause and
Graph display of up to 250 data is permitted.
remedy
If the maximum number of data was exceeded at the time of data add or data move,
there is data that is not displayed. Be cautious.