Invacare® HomeFill® II Oxygen System
1. Prior to each use, inspect product for visible damage.
Refer to the External Examination section. DO NOT use
if any damage is found.
If for any reason, any label becomes illegible or
lost, contact your equipment provider.
2. Examine the cylinder gauge. If the cylinder pressure is
less than 1500 psig (10342 kPa), proceed to STEP 3.
For location of the cylinder gauge, refer to the
cylinder's manual.
If the cylinder pressure is greater than 1500 psig
(10342 kPa), DO NOT attempt to top off this
cylinder. It may not fill.
3. Set the regulator flow dial on the cylinder A to Off.
Refer to the cylinder's manual.
4. Remove the cylinder and connector fillport covers (if
present).
5. Momentarily push DOWN on the outer ring (sleeve) B
of the connector fillport C until GREEN dots D are
visible to reset the connector.
If the outer ring (sleeve) is in the UP position
(GREEN dots not visible), the connector fillport
will not be able to accept the cylinder fillport E.
Pushing DOWN F momentarily will reset the
connector fillport (GREEN dots visible) to accept
the cylinder fillport.
6. Grasp the cylinder in the area behind the cylinder
fillport.
7. Position the cylinder in the compressor cradle G.
22
8. Align the cylinder fillport with the connector fillport.
9. Pull UP on the outer ring (sleeve) of the connector
fillport while pushing DOWN on the cylinder to couple
the cylinder fillport into the connector fillport.
The cylinder is properly connected when an
audible "click" is heard.
4.4 Disconnecting the Cylinder from the
Compressor
1195662-C