Safety Information (Cont.)
of ISO 10993-5:2009 and ISO 10993-10:2010.
It will not cause any potential sensitization or
irritation reaction.
The included adapter is specified as
ME EQUIPMENT.
If you experience discomfort during a
measurement, such as pain in the arm or
other complaints, press
to release the air
immediately from the cuff. Loosen the cuff and
remove it from your arm.
If the cuff pressure reaches 40 kPa / 300
mmHg, the cuff will automatically deflate.
Should the cuff not deflate when pressure
reaches 40 kPa / 300 mmHg, detach the cuff
from the arm and press
to stop inflation.
Before use, make sure the monitor and
accessories function safely and are in proper
working condition. Do not use the monitor or
accessories if they are damaged in any way.
The continuous use of a damaged monitor
10
or damaged accessories may cause injury,
improper results, or serious danger.
Do not wash the cuff in a washing machine or
dishwasher.
The service life of the cuff may vary based on
the frequency of washing, skin condition, and
storage state. The typical service life is 10,000
times used.
We recommend that the performance of the
monitor and cuff should be checked every 2
years and after maintenance and repair, by
retesting at least the requirements in the limits
of the error of the cuff pressure indication and
air leakage (testing at least at 50 mmHg and
200 mmHg).
Please dispose of ACCESSORIES, detachable
parts, and the ME EQUIPMENT according to the
local guidelines.
Safety Information (Cont.)
The manufacturer will make available on
request circuit diagrams, component part lists,
descriptions, calibration instructions, etc., to
assist service personnel in parts repair.
The plug/adapter plug pins insulate the monitor
from the main power supply. Do not position
the monitor in a position where it is difficult
to disconnect from the supply mains to safely
terminate operation of ME equipment.
The person operating this device shall not touch
the output of the batteries/adapter and the
patient simultaneously.
Cleaning: Dust may affect the performance of
the unit. Use a soft cloth to clean every part of
the monitor and cuff before and after use. Do
not use any abrasive or volatile cleaners.
The monitor does not need to be calibrated
within two years of reliable service.
If you have any problems with this monitor,
such as setting up, maintaining, or using, please
contact Customer Support (see page 56). Don't
open or repair the monitor by yourself in the
event of malfunctions. This device must only be
serviced, repaired, and opened by individuals at
authorized sales/service centers.
Please contact Customer Support (page 56) if
any unexpected operation or events occur.
Keep out of reach of infants, young children,
and pets to avoid inhalation or swallowing of
small parts. This is dangerous or even fatal.
Be careful to avoid strangulation due to cables
and hoses, particularly due to excessive length.
At least 30 min are required for ME equipment
to warm from the minimum storage
temperature between uses until it is ready
for intended use. At least 30 min are required
for ME equipment to cool from the maximum
storage temperature between uses until it is
ready for intended use.