EMC Guidance

This ME EQUIPMENT or ME SYSTEM is suitable for
home healthcare environments.
Warning: Don't use near active HF surgical
equipment or the RF shielded room of an ME
system for magnetic resonance imaging, where the
intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or
stacked with other equipment should be avoided
because it could result in improper operation. If
such use is necessary, this equipment and the other
equipment should be observed to verify that they
are operating normally.
Warning: Use of accessories, transducers and
cables other than those specified or provided by
the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and
result in improper operation.
46
Warning: Portable RF communications equipment
(including peripherals such as antenna cables and
external antennas) should be used no closer than 12
in / 30 cm to any part of this equipment, including
cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment
could result.
Technical Description:
1.
All necessary instructions for maintaining
BASIC SAFETY and ESSENTIAL PERFORMANCE
with regard to electromagnetic disturbances for
the excepted service life.
2.
Guidance and manufacturer's declaration -
electromagnetic emissions and immunity.
Table 1
Guidance and Manufacturer's Declaration – Electromagnetic Emission
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Class A
Comply