Table 2
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Immunity Test
IEC 60601-1-2 Test level
Electrostatic
±8 kV contact
discharge (ESD)
±2 kV, ±4kV, ±8 kV,±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV signal input/output
IEC 61000-4-4
100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
Surge
±0.5 kV, ±1 kV,±2 kV common mode
IEC61000-4-5
Voltage dips, short
interruptions and
0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
voltage variations
180°, 225°, 270° and 315°.0 % UT; 1 cycle
on power supply
and 70 % UT; 25/30 cycles; Single phase:
input lines
at 0°.0 % UT; 250/300 cycle
IEC 61000-4-11
48
Table 2
Immunity Test
Compliance Level
±8 kV contact
Power frequency
±2 kV, ±4kV, ±8 kV,
magnetic field
±15 kV air
IEC 61000-4-8
±2 kV for power supply lines
±1 kV signal input/output
Conduced RF
100 kHz repetition frequency
IEC61000-4-6
±0.5 kV, ±1 kV differential
mode
±0.5 kV, ±1 kV,±2 kV common
Radiated RF
mode
IEC61000-4-3
0 % UT; 0,5 cycle. At 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°. 0 % UT; 1 cycle and
NOTE: UT is the a.c. mains voltage prior to application of the test level.
70 % UT; 25/30 cycles; Single
phase: at 0°.
0 % UT; 250/300 cycle
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC 60601-1-2 Test level
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance Level
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio
bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
49