6.5 UUT (UNIT UNDER TEST) SETUP: GENERAL
Screen Instructions (Diagram 15):
UUT setup screeen is used for inputing UUT profiles into the UUT database.
1.
Select an Existing Profile - Go to the UUT ID drop down box, click the down arrow and select
your unit from the list.
2.
Create a New Profile - Tap NEW button on right of page, a dialog box will appear prompting
user to enter new UUT ID. Enter new ID in the box and select UUT to copy if desired from
drop down list below.
Press OK to create the new profile or CANCEL to cancel action.
Note: To automatically setup a UUT using Setra's configurable part number, simply enter the Setra
part number as the new ID, then press "OK". If the number entered is a valid configurable number,
the UUT will be automatically setup.
3.
Enter the UUT's Serial Number in the field provided.
4.
In the Accuracy field, enter the pass/fail accuracy for the UUT.
5.
In the recal Accuracy field, enter an accuracy (greater than the one entered in the Accuracy
field) at which you would like the calibrator to recal the transducer in spite of a passing
result.
Note: Only use with Setra 269. This feature is for automated calibrations, is optional and can
be useful for transducers that have drifted, but are not yet failing, by avoiding a failing result at the
next calibration.
6.
Enter the transducer's lowest and highest pressure in the LOW and HIGH boxes in the Input
column.
7.
Enter the transducers corresponding lowest and highest output in the LOW and HIGH boxes
in the Output column.
8.
Select Test Profile from drop-down list or create a new one (save UUT, go to PROFILE page
and create NEW, return to UUT and select newly created profile for default profile and press
SAVE).
9.
Select Excitation Level from drop down list.
10. Select UUT Type from drop down list, (See Page 28 for UUT Type details.)
11. Save - Once all parameters have been changed, tap the SAVE button at bottom.
button will be 'grayed out' when data has been saved
12. Delete - Select UUT from list of available UUT ID's at the top of the screen. Tap DELETE
button at bottom of the page.
13. More - Press to access remaining UUT setup items.
.
24
Note: Save
Need help?
Do you have a question about the MicroCal and is the answer not in the manual?