I. Set the CH 2 VOLTS/DIV switch to 50 mV.
m. Repeat parts d through i except adjust the " 3
and " 3N" trimmers in parts f and i respectively.
n. Set the CH 2 VOLTS/DIV switch to 0.5 V.
o. Repeat parts d through i except adjust the " 4"
arid "4N " trimmers in parts f and i respectively.
p. Set the Vertical MODS switch to CH 1.
q. Repeat parts b through p for the Channel 1
Attenuators.
r. Disconnect the test equipment from the
instrument.
9. Check Vertical ALT Operation
a. Set:
AC-GND-DC (both)
Vertical MODS
SEC/DIV
Trigger SOURCE
b.
Position the Channel 1 and Channel 2 traces
about two divisions apart using the CH 1 and CH 2
POSITION controls.
c. CHECK—Channel 1 and Channel 2 traces
move across the screen alternately.
d. Set Vertical switch to CHOP.
e. CHECK—Channel 1 and Channel 2 traces
move across the screen at the same time.
10. Adjust CHOP Switch Balance (R140)
a. Set:
SEC/DIV
Trigger MODE
Trigger SOURCE
b. Connect the 10X probe to the rear of R242 on
top of the A l-M ain circuit board and adjust test
oscilloscope for a 4-division display.
C. ADJUST—Chop-Sw Bal (R140) for no trig
gering on chop segments when rotating the Trigger
LEVEL control.
11, Check ADD MODE Operation
b. Position both traces on the center horizontal
graticule line using the CH 1 and CH 2 POSITION
controls. '
c. Set the calibration generator to produce a
50-mV signal.
d. Connect the output of the calibration generator
to both the CH 1
with dual-input coupler.
GND
e. Check that both channels show a 2.5-division
BOTH. NORM
display.
and ALT
0.1 s
f. Set: Vertical MODE switch to ADD.
CH 1
g. CHECK—The resultant display is five divisions
3% (4.85 to 5.15 divisions).
h. Disconnect the test equipment from the
instrument.
12. Adjust High-Frequency Compensation
a. Set:
VOLTS/DIV (both)
Vertical MODE
SEC/DIV
b. Connect the positive-going, fast-rise, square-
wave output from the calibration generator via a
5 0 -0 coaxial cable, a 10X attenuator, and a 5 0 -0
termination to the CH 1 OR X input connector.
1 ms
NORM
c. Set the generator to produce a 1-MHz, 5-
VERT MODE
dtvision display.
d. Set the top of the display to the center hori
zontal graticule Tine using the CH 1 POSITION
control.
e. ADJUST-HF Comp (C242 and C243) for the
5% or less overshoot (0.3 division) on the displayed
signal.
Adjustment Procedure—2205 Service
a. Set:
VOLTS/DIV (both)
AC-GND-DC (both)
Vertical MODE
SEC/DIV
Horizontal MODE
Trigger SOURCE
X input and the CH 2
o r
(C241, C242, C243, and R241)
20 mV
DC
BOTH. NORM,
and ALT
0.5 ms
XI
CH 1
V input
o r
10 mV
C H I
0.2 ms