Parameter Menus - Siemens SITRANS LR560 Operating Instructions Manual

Radar transmitters (foundation fieldbus)
Hide thumbs Also See for SITRANS LR560:
Table of Contents

Advertisement

Commissioning
6.5 Commissioning via handheld programmer
6.5.5.3

Parameter menus

1. Enter PROGRAM
mode
2. Editing in PROGRAM
mode
Note
Navigation
• In Navigation mode, ARROW keys move to the next menu item in the direction of the
arrow.
• For Quick Access to parameters via the handheld programmer, press Home
then enter the menu number, for example: 3.2. Echo Profile.
50
Using local display buttons:
Press RIGHT
Using handheld programmer:
1. Point the programmer at the display from a maximum distance of
300 mm (1 ft).
2. RIGHT
activates PROGRAM mode and opens menu level 1.
3. Mode opens the menu level last displayed in PROGRAM
the last 10 minutes, or menu level 1 if power has been cycled since
then.
To select a listed option:
1. Navigate to the desired parameter.
2. Press RIGHT
to open parameter view.
3. Press RIGHT
again to open Edit mode. The current selection is
highlighted. Scroll to a new selection.
4. Press RIGHT
to accept it.
5. The local display returns to parameter view and displays the new se-
lection.
To change a numeric value:
1. Navigate to the desired parameter.
2. Press RIGHT
to open parameter view.
3. The current value is displayed.
4. Press RIGHT
again to open Edit mode. The current value is high-
lighted. Key in a new value. Press RIGHT
5. The local display returns to parameter view and displays the new se-
lection.
to accept it.