DO NOT USE FOR ORAL SURGERY
BEFORE OPERATING THE PROPHY ANGLE, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS.
SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Indications For Use
Used intraorally by trained dental professionals for drilling and preparation of dental cavities for restoration,
such as fillings, disking, cavity and crown preps, polishing, post and pin drilling and pin setting.
Under normal operating conditions, air is exhausted from the back of the head of this
WARNING
handpiece. If air is directed into an open soft tissue wound or beneath the mucosa
or dermis, injury to the patient could result from air emphysema of air embolism.
This angle has no internal air or water cooling capabilities. External water for
WARNING
cooling maybe required.
Contents
• Troubleshooting
• Features
• Cartridge Replacement
• Installation and Operation
• Replacement Part Numbers
• Safety Rule
• Warranty
• Lubricating
• Cleaning and Sterilization
Features
Rated for use with StarDental Motors and Attachments
Titan Five Star Prophy Angle:
®
StarDental
Titan 3 Part #265803
Controlled Speed Reduces Prophy Paste 'Spin-off'
Field Replaceable Cartridge
Quick, Easy Cleaning
Compatible with Titan 3 and Titan T Systems
Steam Sterilization: 135° C
135°C
Sterilizable up to 135°C
Motor
Accessories
Ratio
5K
MTA
1:1
20K
Torque Multiplier & MTA
4:1
Installation and Operation
Prophy Angle Installation for Titan 3 or Titan T Systems
1. For the Titan 3 or Titan T systems thread angle onto motor-to-angle
adapter, StarDental Titan 3 Part #264073 or Titan T Part # 263970,
by hand. Seat completely.
When angle does not seat on adapter, gears are not meshed correctly. Remove
CAUTION
angle; spin motor-to-angle adapter drive gear and repeat step one.
2. Fit angle disconnect wrench, StarDental Part
#203048, furnished with every angle, to mating
indentations on angle neck; tighten securely.
Prophy Cup Installation
The prophy angle uses either a screw-on or a
snap-on type polishing cup. When using a
snap-on type cup, use the screw-to-button
adapter, StarDental Part #258950.
Thread the screw-to-button adapter, or polishing
cup, with a clockwise rotation until snug into the
hollow cartridge shaft of the prophy angle.
Operate the Titan motor in the forward direction only. If the
CAUTION
Titan motor is operated in reverse direction, the prophy cup
will separate from the motor.
Prophy Cup Removal
Unthread the adapter, or polishing cup, counterclockwise. If the
screw-to-button adapter is difficult to disengage, remove the prophy cup,
and use a small screwdriver to loosen the screw-to-button adapter.
SAFETY
INSTRUCTIONS
PLEASE OBSERVE SAFETY INSTRUCTIONS
Before use, verify concentric rotation by operating the handpiece outside the oral cavity in a safe place.
Lubricating
1. Lubricate with DentaLube® II, StarDental® Part #262539 once a day or before each sterilization
procedure.
2. Remove prophy cup.
3. Place a drop of DentaLube II down the center of the hollow cartridge shaft (allow a few seconds
for the lubricant to drain down the shaft). (See illustration)
4. Place a drop of DentaLube II at the base of the angle. Allow a few seconds for the lubricant
to drain and reach the bronze bushing. (See illustration)
Red Tip Applicator
Max Allowable RPM
5,000
5,000
Cleaning & Sterilization
It is the responsibility of the user to use only FDA compliant sterilizing equipment and
Motor-to-Angle
CAUTION
accessories that are capable of meeting the manufacturers recommendations below.
Adapter Drive Gear
Do not exceed 135°C
Do not use: • dry heat sterilizer
CAUTION
Limits on
Repeated processing has minimal effect on these angles. End of life is
normally determined by wear and damage due to use.
Reprocessing
INSTRUCTIONS
Titan
Mating Indentations
Prophy Angle
The angle must be cleaned and sterilized before first use and between patient use.
WARNING
Containment &
It is recommended that angles are reprocessed as soon as is reasonably
Disconnect
transportation:
practical following use.
Wrench
Preparation
Wear puncture resistant gloves to prepare the prophy angle for sterilization.
for cleaning:
Remove prophy cup or screw-to-button adapter from prophy angle. Remove
Motor-
angel from motor-to-angle adaptor.
to-Angle
Washer-disinfectors are not recommended.
Cleaning
Adapter
Automated:
1. Clean and inspect once a day.
Cleaning
Manual:
2. Clean all external surfaces with
Screw-to-
3. Use the plastic wrench, StarDental
Button
Adapter
4. Remove the cartridge and inspect as
5. If prophy paste
6. If prophy paste
Red
Tip Applicator
NOTICE
Do not ultrasonically clean any other components.
• cold sterilizers • ultrasonic cleaners
•immersion techniques
• sterilizer bags containing disinfectants
• detergents
• surface wipes
•surface sprays
• chemical disinfectors
• hand soaps
• surgical milk
Unthread End Cap in
isopropyl alcohol and a lint-free cloth
Clockwise Direction
or brush.
Part #258953, to unthread the end
cap by turning clockwise. See
illustration:
Plastic
Wrench
follows:
New Style:
Old Style:
is not found on
-c 34 o and L
-No Markings on End Cap
t he cartridge,
Marked On End Cap
internal cleaning
- Unthread End Cap in
- Unthread End Cap in
CounterClockwise Direction
is not necessary.
Clockwise Direction
Simply wipe off
wit h lint-free
cloth.
i s f o u n d o n
t he cartridge,
cl e a n o f f t h e
paste manually. The gear
Housing
Gear Assembly
assembly and end cap may be
ultrasonically cleaned.
Bronze Bushing
Cartridge
O-ring
End Cap

Advertisement

loading

Summary of Contents for DentalEZ StarDental Titan 3 Five Star

  • Page 1 Installation and Operation Cleaning & Sterilization Prophy Angle Installation for Titan 3 or Titan T Systems It is the responsibility of the user to use only FDA compliant sterilizing equipment and Motor-to-Angle CAUTION 1. For the Titan 3 or Titan T systems thread angle onto motor-to-angle accessories that are capable of meeting the manufacturers recommendations below.