Settings and Measurement
Lower - Level- Upper .......... Specifies the alarm level(s) for the conditions set in Mode.
Output ................................. Specifies the output channels for alarm generation (4 channels).
Pulse
Sets the alarms for pulse input settings. These conditions can be set when they have been enabled
in the AMP settings.
Mode ................................... Sets the same conditions as for "Condition Settings". Please refer to
Lower - Level- Upper .......... Specifies the alarm level(s) for the conditions set in Mode.
3-18
L: An alarm is generated when the signal input falls to (or falls below)
the specified alarm level.
Alarm generated
CH.1
Win In: Specifies the upper and lower alarm limits for each channel.
When the signal level goes within (or is within) both limits, an
alarm is generated.
CH.1
Win Out: Specifies the upper and lower alarm limits for each channel.
When the signal level goes outside (or is outside) both limits,
an alarm is generated.
CH.1
L, H: Input a numeric value.
Win In, Win Out: Input numeric values for the upper and lower levels.
"Condition Settings" in the previous section.
L, H: Input a numeric value
Alarm generated
Alarm generated