Settings and Measurement
Lower - EU - Upper Unit ..... Sets the EU function's conversion value and unit. If the ENTER key is
3-10
pressed here, the following window is displayed.
(a) Meas. Value
(e) Select
(d) Unit
CHECKPOINT
The Scaling operation is calculated using a ratio of the Meas. Value or
EU Output Value settings. If a ratio value that the GL400 cannot
process is specified, the message below appears.
Out of input range
[ENTER] Apply
If this message appears, follow the instructions by reducing the
number of digits to be output by one, or leaving the number of digits
as is and changing the EU value.
(a) Meas. Value
Specifies the numeric value to be converted. Set two points, the
Upper and Lower parameters.
(b) EU Value
Specifies output after conversion. Set two points, the Upper and
Lower parameters.
(c) Dec pt
This parameter specifies the decimal point position of the numeral
to be specified as the EU value(s).
(d) Unit
Selects the converted unit, which can be specified as a user-
defined character string consisting of alphanumerics. The Unit
parameter can also be specified by selecting the Select Unit
setting.
(e) Select
Selects the type of engineering unit.
(f) Choose
Selects the converted unit. The Unit displayed here is the type of
unit selected by the Select setting.
To specify a unit that is not displayed here, specify a user-defined
character string as the Unit setting. Moreover, the setting specified
here is displayed as the Unit setting.
Setting Example
Upper Value
Specified Value
+2.5000
EU Value
+10.000
+5 V
CH.1 10V
–5 V
(b) EU Value
(c) Dec pt
(f) Choose
Lower Value
–2.5000
–10.000
+20.00 rpm
CH.1 Scaling 1
–20.00 rpm