To open the standard click stimulus file, click File, click Open, browse to:
b.
C:\TDT\BioSigRZ\SIG\Click.sig
Then click Open.
c.
On the main toolbar, click Modify, and then click Signal.
d.
In the Signal Parameters dialog box, enter the voltage and the corresponding dB level in
the Calibration area, and then click OK.
5.
To save the change, click File then Save.
This will save the change to the stimulus used by BioSigRZ's standard click configuration.
6.
Run the stimulus file in BioSigRZ without a calibration (.tcf) file applied to the stimulus. Your
stimulus is already calibrated, and adding a .tcf file will cause the calibration to be incorrect.
Calibration | 45