8 . M A I N M E N U
8.3.6 Questionnaire
To enable or disable the questionnaire, select the Analyser Settings option from the main menu, then in the Analyser Settings screen,
select Questionnaire.
In the Questionnaire screen, the first option enables the questionnaire set and the second option disables it. Use the five way navigation key to
select the required settings. Press OK to save the settings.
8.3.7
Automatic shutdown
The SoToxa™ Mobile Analyser can be set to automatically shutdown after a period of inactivity to save power. The automatic shutdown
function can be set to on or off and the time period before shutdown can be selected.
Select Analyser Settings from the main menu, then in the Analyser Settings screen, select Auto Shutdown. In the Auto Shutdown screen, use
the five way navigation key to highlight the required setting and press OK to save the setting.
The analyser will not automatically shutdown while a test is in progress or if a SoToxa™ Test Cartridge is inserted.
8.3.8 Periodic QC
The SoToxa™ Mobile Analyser can be set to enforce QC testing at a defined time intervals.
Select Analyser Settings from the main menu, then in the Analyser Settings screen, select Periodic QC. In the Periodic QC screen, use the five
way navigation key to highlight the required setting and press OK to save the setting.