Review 07/20
02
INSTALLATION
02.2

Anchorage points

Before putting the stretcher support into operation, ensure that each and every anchorage points are pla-
ced and clasped.
ANCHORE FOR
PB-720 · PB-725
(8 anchorage points)
LIMIT AMBULANCE FLOOR
441mm
MAX.
185
256
102
IMPORTANT FOR INSTALLATION
- Screws to be attached on the floor should be M12 according to norm DIN-7991 (8 units)
- Before drilling the ambulance floor, verify that the lateral movement of the stretcher su-
pport is placed at 0 position
- Install it upon the limits indicated on the drawing and make sure to fix the set on the rein-
forced chassis are
- The electric installation of the set must be completed with the electrical system of the
ambulance disconnected
USE SCREWS DIN-7991 - M12 - 10.9
12,5 (X8)
1274
2042
LIMIT AMBULANCE WALL
102
180mm
MIN.
13