Ametek Reichert Tono-Vera Vet Instructions For Use Manual page 32

Tonometer
Table of Contents

Advertisement

Guidance Tables
Table 9 – Guidance and Manufacturer's Declaration
Electromagnetic Immunity
Immunity to Proximity Fields from RF Wireless Communications Equipment
Guidance and Manufacturer's Declaration - Electronic Immunity
Tono-Vera Vet is intended for use in the electromagnetic environment as specified below related to proximity fields from RF wireless communications equipment.
Immunity Test
IEC 60601 Test Level
Test
Frequency
Band (MHz)
(MHz)
385
380-390
450
430-470
710
745
704-787
780
810
Radiated RF IEC
61000-4-3
870
800-960
930
1720
1845
1700-1990
1970
2450
2400-2570
5240
5500
5100-5800
5785
(continued)
Service (MHz)
Modulation
Pulse Modulation
TETRA 400
18 Hz
FM ±5 kHz
GMR 460, FRS 460
deviation 1 kHs
sine
Pulse Modulation
LTE Band 13, 17
217 Hz
GSM 800/900, TETRA 800,
Pulse Modulation
iDEN 820, CDMA 850, LTE
18 Hz
Band 1, 3, 4, 25; UMTS
GSM 1800; CDMA 1900;
Pulse Modulation
GSM 1900; DECT; LTE Band
217 Hz
1, 3, 4, 25; UMTS
Bluetooth WLAN, 802.11
Pulse Modulation
b/g/n, RFID 2450, LTE
217 Hz
Band 7
Pulse Modulation
WLAN 802.11 a/n
217 Hz
Immunity
Maximum
Distance (m)
Test Level
Power (W)
(V/m)
1.8
0.3
27
2
0.3
28
0.2
0.3
9
2
0.3
28
2
0.3
28
2
0.3
28
0.2
0.3
9
Electromagnetic
Compliance Level
Environment
- Guidance
Compliance Level
27 V/m at 0.3 m
28 V/m at 0.3 m
d = 6/E √P
9 V/m at 0.3 m
where
d = Minimum separation
distance in meters
28 V/m at 0.3 m
E = Immunity test level
in V/m
P = Maximum power in
Watts (W)
28 V/m at 0.3 m
28 V/m at 0.3 m
9 V/m at 0.3 m
16308-101 Rev. A • 32