Cleaning and Maintenance

Cleaning Tono-Vera Vet

CAUTION: Do not immerse Tono-Vera Vet in fluids. This will cause
damage to the electronics and void the warranty.
WARNING: Do not spray, pour, or spill liquid onto Tono-Vera Vet, its
accessories, connectors, switches, or openings in the body. Remove any
liquid appearing on the surface of Tono-Vera Vet immediately.
The outer surfaces of Tono-Vera Vet and Tono-Vera Vet Base, including
the charging contacts, may be safely cleaned with the following:
7.5% hydrogen peroxide and water solution
5.25% bleach and water solution (sodium hypochlorite at least 5000
parts per million (ppm) diluted at 1/4 unit volume 5.25% bleach to
2.25 unit by volume distilled water)
70% isopropyl alcohol
Quaternary ammonium compounds
Press and hold the center Multifunction Button for three seconds
1
to turn the power off.
Dampen a soft cloth with one of the approved solutions,
2
or use one of the approved products.
Lightly wipe the surfaces.
3
Remove residual fluid using a soft, dry cloth.
4
(continued)

Battery

Replace Tono-Vera Vet batteries when needed. Only replace with new
AA batteries or the Reichert Tono-Vera Vet Lithium-Ion Rechargeable
Battery Pack. Energizer® or Duracell® batteries are recommended
to be used with this device. Alternative brands may impact device
measurement longevity.
NOTE: Only use new batteries. Never mix new and old batteries together.

Storage

If the device is to be stored for an extended period, remove the AA
batteries or the rechargeable battery pack to avoid possible damage
to the device due to battery leakage. The device may be stored on the
Tono-Vera Vet Base or in the Tono-Vera Vet Carry Case.

Disposal

Tono-Vera Vet, Ocu-Dot Probes, and Tono-Vera Vet Lithium-Ion Cartridge
do not generate any environmentally hazardous residues. At the end of
its product service life, follow your local laws and ordinances regarding
the proper disposal of the device.