Ocu-Dot Probe Installation

Pull off the Tono-Vera Vet Protective Probe Chamber Cover from the
1
Tono-Vera Vet Probe Chamber. Take a new Ocu-Dot Probe Tube from
the box and remove the cap.
Insert the open end of the Ocu-Dot Probe Tube into the Probe Chamber.
2
Rotate Tono-Vera Vet up so the Ocu-Dot Probe can slide into the Probe
3
Chamber.
To remove the Ocu-Dot Probe, press and hold the center Multifunction
4
Button for three seconds to turn the device off and eject the Ocu-Dot
Probe. It is recommended to eject the used Ocu-Dot Probe directly into
the Probe Tube for disposal.
You may store the empty Ocu-Dot Probe Tube in recess on base.
5
If Tono-Vera Vet is on, the tonometer will automatically secure the Ocu-Dot
Probe. If it is not, press any button to turn the device on and secure the
Ocu-Dot Probe.
NOTE: The Probe Tube can be used to dispose of the probe, which may be
placed in a sharps container to prevent any injury.
WARNING: To prevent contamination, do not touch the bare Ocu-Dot Probe.
Do not use a probe if it touches a non-sanitized surface like a table or a
floor. Properly dispose of a touched or dropped probe (e.g. in a container
for disposable needles).
WARNING: Do not use an Ocu-Dot Probe on more than one patient to
prevent cross contamination.
WARNING: When Tono-Vera Vet turns off, the probe will be ejected.
(continued)
1
2
3
5
Store Ocu-Dot Probe
Tube in recess.
16308-101 Rev. A • 13