Measuring

If Measure Mode is set to ActiView Auto Measure, measurements will be
1
taken automatically when Tono-Vera Vet is aligned correctly. If Measure
Mode is set to ActiView Manual Measure or Manual Measure, press the
blue Manual Measure Button to measure upon correct alignment.
NOTE: The blue Manual Measure Button can be pressed at any time if the
operator wishes to override the auto measurement system. However, this
may result in less reliable measurements. Measurements will only be
taken if the device is within the ±15º probe angle range.
The measurement dots indicate the number of and status of
2
measurements needed to obtain the Final IOP Result.
Each empty white dot indicates a required measurement. A solid blue
dot indicates a successful measurement. A solid yellow dot indicates an
unsuccessful measurement which must be retaken.
When Measurements are set to 1 (Quick), only one meaurement will
be needed to obtain a high-confidence IOP result. When set to 3+,
measurements will be taken until a high-confidence IOP result is obtained,
a minimum of three measurements and a maximum of six. If more than
three measurements are required, an empty gray dot will turn white,
indicating another measurement is required.
NOTE: When measurements are completed, the IOP will be displayed in
a large colored ring on the screen. Refer to the Understanding Results
section of this manual for detailed information.
After measuring the first eye, the device will automatically select the next
3
eye to measure.
WARNING: If the device takes a measurement while the patient blinks, the
probe could potentially become caught by the patient's eyelashes. If this occurs,
simply dispose of the probe and begin the examination again with a new probe.
CAUTION: Do not place device on a surface with the Ocu-Dot Probe or
Probe Chamber facing down. This may damage the device and Ocu-Dot
Probe. Place device on its side if the device must be laid down on a surface.
(continued)
1
Manual
Measure
Button
2
Number of
Measurements
12.
5
OD IOP
Clear
Select and Clear
OD/OS Buttons
3
14.
5
16308-101 Rev. A • 17