Guidance Tables
Table 201 – Guidance and Manufacturer's Declaration
Electromagnetic Emissions
All Medical Electrical Equipment and Medical Electrical Systems
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Tono-Vera Vet is intended for use in the electromagnetic environment specified below. The customer or user of Tono-Vera Vet should ensure that it is used in such an environment.
EMISSIONS TEST
Conducted and Radiated RF Emissions
CISPR 11
Conducted and Radiated RF Emissions
CISPR 11
Harmonic Distortion
IEC 61000-3-2
Voltage Fluctuations and Flicker
IEC 61000-3-3
(continued)
ELECTROMAGNETIC ENVIRONMENT
COMPLIANCE
- GUIDANCE -
Tono-Vera Vet uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any
Group 1
interference in nearby electronic equipment.
Class B
Tono-Vera Vet is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-
voltage power supply network that supplies building for domestic power.
Class A
Complies
16308-101 Rev. A • 28