XTP601 User's Manual
Register Type C: Display Parameters
15
14
13
12
Display Brightness
r/w
r/w
r/w
r/w
0–100 in 10% steps
Register Type D: Units
15
14
13
12
r/w
r/w
r/w
r/w
Bits
0, 1
2, 3
4
5
11,12,13
Register Type F: -2000.0 to +2000.0
15
14
13
12
r/w
r/w
r/w
r/w
Range
= 0 to 40000 represents -2000.0 to +2000.0
Conversion:
(RegValue – 20000)/10.0
For external sensor values
Dew point:
-100/+20 °C, -148.0/+68.0°C, 173.0/293.0 K
Tempr:
-50.0/+100.0 °C, -58.0/+212.0 °F, 223.0/373.0 K
Pressure:
0.0/44.1 psia, 0.0/3.0 barA, 0.0/304.0 kpa
Register Type G: -200.00 to +200.00
15
14
13
12
r/w
r/w
r/w
r/w
Range
= 0 to 40000 represents -200.00 to +200.00
Conversion:
(RegValue – 20000)/100.00
Michell Instruments
11
10
9
8
r/w
r/w
r/w
r/w
11
10
9
8
r/w
r/w
r/w
r/w
HEX
0003
000C
Ext press unit, 00 = psia, 01=bara, 10=kPa
0010
0020
Ext Sensor Parameter (000=none, 001=dewp, 010=tempr,
3800
11
10
9
8
r/w
r/w
r/w
r/w
11
10
9
8
r/w
r/w
r/w
r/w
7
6
5
4
Display Contrast
r/w
r/w
r/w
r/w
0–100 in 10% steps
7
6
5
4
r/w
r/w
r/w
r/w
Meaning (binary)
00=°C, 01=°F, 10=K
Field cal type, 0=1 gas (offset), 1=2 gas
Date format 0=Non US, 1=US
011=press, 100=other)
7
6
5
4
r/w
r/w
r/w
r/w
7
6
5
4
r/w
r/w
r/w
r/w
APPENDIX B
3
2
1
0
r/w
r/w
r/w
r/w
3
2
1
0
r/w
r/w
r/w
r/w
3
2
1
0
r/w
r/w
r/w
r/w
3
2
1
0
r/w
r/w
r/w
r/w
43
Need help?
Do you have a question about the XTP601 Series and is the answer not in the manual?