INTRODUCTION
1.1
Features
•
The XTP601 can be supplied meeting the requirements of IEC 61508
SIL2, allowing the user to integrate the analyzer into their functional
safety system.
•
There are 2 versions of the XTP601 available:
General Purpose Analyzer with touch screen (capacitive buttons)
Hazardous Area version of above with flame arrestors
•
The XTP601 is calibrated in a specific background gas to match
customer's requirements. This is displayed on the front screen of the
analyzer or via the Application Software.
•
The XTP601 provides 2 off 4-20 mA analog output signals that are
proportional to the oxygen concentration. The primary 4-20 mA output
is locked on the calibrated range of the instrument. The secondary
output is user-selectable within the calibrated range.
•
Modbus RTU over RS485 protocol serial communications are provided
as standard.
•
Accuracy of 1% of reading or 0.02% O
suppressed ranges ±0.2% O
•
Zero and span stability of 0.25% of span per month.
•
The innovative Thermo-Paramagnetic sensor design used in the XTP601
Oxygen Analyzer has no moving parts. This makes it less sensitive
to vibration and movement and highly resistant to drift over a long
period of time, compared to other sensing technologies, and it offers
increased protection against contamination.
•
The robust and weather-proof enclosure (IP66 rated) allows the XTP601
to be installed on-site, at the point of measurement. It is also available
with an explosion-proof casing.
•
Low cost of ownership due to minimal maintenance. For installation
and commissioning, the casing unscrews to allow for easy access to the
electrical connections.
•
The XTP601 is certified to ATEX, IECEX, cQPSus, INMETRO & TC TR Ex
for use in hazardous areas.
•
2 x single pole change-over relay alarms for oxygen concentration,
supplied as standard. Can be configured as OFF, LOW or HIGH.
2
, whichever is greater. (Zero
2
).
2
97313 Issue 8.1, November 2020
XTP601 User's Manual
Need help?
Do you have a question about the XTP601 Series and is the answer not in the manual?