XA01006T
Type
TMT82
TMT84, TMT85
Electrical
connection data
Type
Type of
protection
TMT82
Ex ia IIIC T85°C...
T120°C Dc
TMT84, TMT85
Ex ia IIIC T85°C...
T120°C Dc
TMT82
Ex tc IIIC T85°C...
T105°C Dc
Ex tc IIIC T105°C
Dc
Ex tc IIIC T120°C
1)
Dc
Ex nA IIC T6...T4
Gc
TMT84, TMT85
Ex tc IIIC T85°C...
T105°C Dc
Ex tc IIIC T105°C
Dc
Ex tc IIIC T120°C
1)
Dc
Ex nA IIC T6...T4
Gc
1)
allowable maximumsurface temperature on the housing for transmitters without provided terminal head
16
Type of protection
Ex nA IIC T6...T4 Gc
Ex nA IIC T6...T4 Gc
Power supply
(terminals 1+
and 2-)
U
≤ 30 V
i
DC
I
≤ 130 mA
i
C
= negligible
i
small
L
= negligible
i
small
U
≤ 17.5 V
i
DC
I
≤ 500 mA
i
C
= 5 nF
i
L
= negligible
i
small
U
=
b
11 to 42 V
DC
Output:
4 to 20 mA
U
= 9 to 32 V
b
Output:
FOUNDATION
Fieldbus™
PROFIBUS PA®
Current consum.
≤ 11 mA
Ambient temperature
–40 °C ≤ Ta ≤ +58 °C
–40 °C ≤ Ta ≤ +75 °C
–40 °C ≤ Ta ≤ +85 °C
–40 °C ≤ Ta ≤ +55 °C
–40 °C ≤ Ta ≤ +70 °C
–40 °C ≤ Ta ≤ +85 °C
Sensor circuit
Max. connection values
(terminals 3 to
7)
U
≤ 7.6 V
Ex ia IIIC/IIIB/IIIA L
o
DC
I
≤ 13 mA
o
P
≤ 24.7 mW
o
U
≤ 7.2 V
Ex ia IIIC/IIIB/IIIA L
o
DC
I
≤ 25.9 mA
o
P
≤ 46.7 mW
o
DC
iTEMP TMT82, TMT84, TMT85
Temperature class
T6
T5
T4
T6
T5
T4
= 50 mH
o
C
= 4.5 µF
o
= 50 mH
o
C
= 4.6 µF
o
Endress+Hauser