Opc Power-Up Test - Aqua Lung PHODS Technical Manual

Portable helicopter oxygen delivery system
Hide thumbs Also See for PHODS:
Table of Contents

Advertisement

14

OPC POWER-UP TEST

Power On/Off & Mode
Selector Knob
Low Battery
Oxygen Flow
Warning LED
Indicator LED
1. Insert the 90° fitting on the OPC inlet hose (36) into the
CPC connector (9) on the regulator. Insert the blue tipped
end of the outlet hose (35) into the 6mm connector (39) on
the OPC. Attach the coiled hose (33) to the CPC connector
on the outlet hose (35). Insert the coiled hose (33) into the
90° fitting on the index tube of the cannula (32) or facemask.
2. Turn the mode-selector knob to the ON position. A
successful Power-Up Test consists of the following: Both LEDs
flash 6 times, audible alarm beeps 3 times and a single puff
of oxygen should be emitted from the cannula via the OPC.
If during the Power-Up Test the battery LED displays continuous
single or double flashes, this indicates a low battery condition.
Replace the batteries immediately before continuing.
NOTE: The OPC will not deliver an oxygen pulse below
8,000 ft. PA while in ON mode other than during the
Power-Up Test.
Outlet Hose
Inlet Hose
Static Ports
(DO NOT COVER)
3. Turn the mode-selector knob to the R/M position. This will allow
the unit to provide a puff of oxygen while breathing from the unit.
4. Turn the mode-selector knob to the F20 position. This will allow
the unit to provide a puff of oxygen while breathing from the unit.
5. Upon successful completion of the OPC Power-Up
Test, turn the mode-selector knob to the OFF position.
WARNING:
If the OPC fails the function test,
Table 2: Troubleshooting Guide OPC section of this
manual to determine its possible cause.
Providing PHODS has passed all testing and the cylinder has
been filled to 2100 psi / 145 BAR, PHODS is ready for use.
PHODS Technical Manual
refer to

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents