Table of Contents

Advertisement

i300
Dive Computer Owner's Manual
© Aqua Lung International, Inc. (2015)
Doc. 12-7823-r06 (12/10/15)

Advertisement

Table of Contents
loading

Summary of Contents for Aqua Lung i300

  • Page 1 Dive Computer Owner's Manual © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 2: Notices

    The program within the i300 simulates the absorption of inert gases into the body by using a mathematical model. This model is merely a way to apply a limited set of data to a large range of experiences. The i300 dive computer model is based upon the latest research and experi- ments in decompression theory.
  • Page 3: Warnings

    • If you exceed certain limits, the i300 will not be able to help you get safely back to the surface. These situations exceed tested limits and can result in loss of some functions for 24 hours after the dive in which a violation occurred.
  • Page 4: Table Of Contents

    6. CF (Conservative Factor) 2. dEPtH AL (ALARM) 1-3 7. LIGHt (BACKLIGHT) DURATION SET MODE 8. SR (SAMPLE RATE) SHARED SETTINGS SET TIME INITIATING A DIVE SET MODE FREE DIVE MAIN HISTORY ALT 1 © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 5 SERVICE MODULE REMOVAL FROM A BOOT BATTERY REPLACEMENT RETURNING THE MODULE TO A BOOT ALTITUDE SENSING AND ADJUSTMENT TECHNICAL DATA  NO DECO TIME LIMITS OXYGEN EXPOSURE LIMITS ALTITUDE LEVELS SPECIFICATIONS ABBREVIATIONS/TERMS © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 6: Getting Started

    GETTING STARTED © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 7: Basics

    Dive, Gauge, and Free Mode. Though the i300 is easy to use, you will get the most out of your new i300 if you take some time to familiarize yourself with its displays and operation.
  • Page 8: Display Icons

    Dive Time or # Deep Stop Time To Surface Safety Stop Temperature Deco Stop Triggered Fraction of Oxygen Low Battery Surface Time Descend, Ascend, or Stop Partial Pressure of Oxygen Value is Maximum © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 9: Buttons

    GETTING STARTED BUTTONS The i300 utilizes 2 control buttons called the ADV (Advance) and SEL (Select) buttons. They allow you to select mode options and access specific information. They are also used to enter settings, activate the backlight, and acknowledge the audible alarm. Throughout this manual they will be referred to as the ADV and SEL buttons.
  • Page 10: Button Functions

    BUTTON FUNCTIONS ACTION BUTTON FUNCTION Press Button • to activate the i300 • to access Alt screens • to advance or step through menus • to toggle or change setpoints • to activate the backlight • to scroll quickly, changing setpoints •...
  • Page 11: Dive Features

    DIVE FEATURES © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 12: Dtr (Dive Time Remaining)

    DTR (DIVE TIME REMAINING) The i300 constantly monitors No Deco status and O2 Accumulation, and will display whichever Time is the least amount available as DTR on the No Deco Dive Main screen. The Time being displayed will be identified by the NO DECO or O2 MIN icons.
  • Page 13: Asc Bar Graph

    DS (DEEP STOP) When the DS selection is set to ON, it will trigger after descending deeper than 80 ft (24 m). The i300 then calcu- lates (continually updating) a Stop Depth equal to ½ the Max Depth.
  • Page 14: Ss (Safety Stop)

    • The battery icon appears solid upon entry into Surface Mode. Alarm Level • The i300 functions continue but the backlight is disabled. • The Battery icon appears flashing. 5 seconds after entering Surface Mode the i300 will shut down. © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 15: Audible Alarm

    • Ascent above a Deco Stop for more than 5 minutes. • Deco requires a Stop Depth deeper than 60 ft (18 m) or deeper. • On the Surface during a Conditional Violation. © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 16: Dive Surface Mode

    DIVE SURFACE MODE © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 17: On The Surface Before A Dive

    To view i300 logs, change settings, or switch modes you must navigate through the Surf Main Menu. Enter the menu by pressing the ADV button. When you reach the end of the menu the i300 will return to the Dive Surface Main screen.
  • Page 18: Fly/Sat (Desat)

    (hr:min), which will generally occur prior to the FLY countdown reaching 0:00 (hr:min), the SAT time is to remain on the screen as 0:00 until the FLY counter shuts the i300 off, 24 hours after the last dive. NO PREVIOUS DIVES...
  • Page 19: Plan

    DIVE, time dive GAU for GAUGE dive, started VIO for a violation dive date dive # of dive next item in access next dive enter dive Main Menu recorded record © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 20: Set Gas

    SET GAS LEAD-IN SET FO SET PO ALARM Alarm setting current Alarm setting setting next item in access adjust FO save adjust PO save Main Menu Set FO setting setting setting setting © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 21: Set Al (Alarms)

    2. dEPTH AL (Audible Alarms) The Depth Alarm feature allows you to set a maximum depth alarm. SET DEPTH AL or 30 - 330 ft (10 - 100 m) to change save setting setting © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 22: Edt Al (Elapsed Dive Time Alarm)

    This feature allows you to set an alarm to go off with a designated reserve of dive time remaining. DTR AL DTR AL TRIGGERED OFF or blank flashing if switched on setpoint 5 - 20 min blank if OFF to change save setting setting © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 23: Set Util (Utilities)

    2. H2O ACT (Water Activation) The H2O ACT feature allows you to turn OFF water contact activation. WARNING: With H2O ACT turned OFF, you MUST remember to manually activate the i300 before any dive. SET H2O ACT ON or OFF...
  • Page 24: Units (Imp/Met)

    SET STOP DEPTH ON or OFF 3 or 5 min (10, 15; 20 ft) or (4, 4, 5; 6 m) toggle save toggle save advance save setting setting setting setting setting setting © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 25: Cf (Conservative Factor)

    NOTE: The Backlight utilizes an ambient light sensor. If there is sufficient light, the back- light will not turn on regardless of this setting. SET LIGHT OFF, 5 sec; 10 sec advance save setting setting © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 26: Sr (Sample Rate)

    Download function stores much larger files for each dive. Depending on the chosen set- tings and dive durations, it is possible to see dives stored in the i300's onboard Log that have already been overwritten in the PC Download Partition. Choosing a longer Sample Rate interval will consume less memory per dive.
  • Page 27: Set Mode

    Set Mode allows you to choose between diVE (standard recreational dive), GAUGE, and FrEE (free diving) modes of operation. NOTE: The i300 will be locked in Gauge mode for 24 hours after surfacing from any Gauge or Violation Dive. Otherwise, you may switch modes freely while in any Surface Mode.
  • Page 28: Sn (Serial Number)

    DIVE SURFACE MODE SN (SERIAL NUMBER) Information displayed on the Serial Number screen should be recorded and kept with your sales receipt; it will be required in the event that your i300 requires factory service. SERIAL NUMBER revision # serial #...
  • Page 29: Dive Operation

    DIVE OPERATION © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 30: Initiating A Dive

    DIVE OPERATION INITIATING A DIVE With the i300 activated, a dive will commence upon descending to 5 ft (1.5 m) for at least 5 seconds. Below is a diagram to help you navigate Dive Mode functions. NO DECO DIVE MAIN...
  • Page 31: Dive Alt 1

    DIVE ALT 1 This screen simply tells you the current time of day and ambient temperature. DIVE ALT 2 The ALT 2 screen displays information pertaining to nitrox; it is bypassed if the i300 is set for air. saturation 0 - 100...
  • Page 32: Safety Stop Main

    SURFACING Upon ascending to 3 ft (0.9 m) the i300 transitions to Dive Surface mode. For the first 10 minutes after a dive the i300 will continue to display the maximum depth and elapsed dive time. Once the surface time reaches 10 min- utes the i300 will display the standard Dive Surface screen.
  • Page 33: Deco Entry

    > Once the penalty time is worked off, and off gassing credit begins, required Deco Stop Depths and Time will de- crease toward zero. The N2 Bar Graph will recede into the No Deco zone, and operation will revert to No Deco mode. down arrow flashing N2 Bar Graph flashing © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 34: Dv 1 (Delayed Violation 1)

    Stop icon will be on solid. > If the DV1 status is ignored, the i300 will enter DV1 Surface mode for 5 minutes upon surfacing from the dive. VIO (Violation), Down Arrow icon, and SURF icon will be flashing. After 5 minutes on the surface in DV1 mode, the unit will enter VGM (Violation Gauge Mode).
  • Page 35: Dv 3 (Delayed Violation 3)

    * The maximum functional depth {Dive/Free Mode 330 FT (100 M), Gauge Mode 399 FT (120 M)} is the depth at which the i300 can properly perform calculations or provide accurate display information.
  • Page 36: High Po

    (1.60) with icon will alternate with STOP message once each minute*. on for 10 seconds, Deco Stop Depth/Time on for 50 seconds until PO decreases below 1.60, then PO will not be dis- played. alternate with STOP © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 37: High O2 Sat (Oxygen Saturation)

    MAX O2 SAT (100% O2 SAT) will flash (in place of max depth and TTS) until on the surface. DURING AUDIBLE ALARM AFTER AUDIBLE ALARM % O2 SAT required flashing stop depth required required required flashing stop depth stop time stop time © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 38: Alarm On Surface

    • Access to Dive ALT screens is allowed during the first 10 minutes, then access to the Dive Surface Menu is allowed. SURFACE < 10 MIN, SURFACE > 10 MIN, 100% O2 SAT VGM TRIGGERED % O2 SAT alternating with flashing % O2 SAT © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 39: Gauge Mode

    GAUGE MODE © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 40: On The Surface Before A Dive

    To view i300 logs, change settings, or switch modes you must navigate through the Surf Main Menu. Enter the menu by pressing the ADV button. When you reach the end of the menu the i300 will return to the Dive Surface Main screen.
  • Page 41: Initiating A Dive

    GAUGE MODE INITIATING A DIVE With the i300 activated, a Gauge dive will commence upon descending to 5 ft (1.5 m) for at least 5 seconds. Be- low is a diagram to help you navigate Gauge Dive Mode functions. GAUGE DIVE MAIN...
  • Page 42: Dv 3 (Delayed Violation 3)

    Too Deep. *The maximum functional depth {Dive/Free Mode 330 FT (100 M), Gauge Mode 399 FT (120 M)} is the depth at which the i300 can properly perform calculations or provide accurate display information. Refer to the Specifications in the back.
  • Page 43: Free Mode

    FREE MODE © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 44: Free Dive Mode Details

    No Deco Dive Time Remaining are carried over from one operating mode to the other, which permits the user to maintain awareness of nitrogen absorption and off-gassing status. • The mathematical models currently used in the i300 are based on no decompression/decompression multilevel repetitive dive schedules.
  • Page 45: On The Surface Before A Dive

    To view ALT screens, change settings, or switch modes you must navigate through the Surf Main Menu. Enter the menu by pressing the ADV button. When you reach the end of the menu the i300 will return to the Free Surface Main screen.
  • Page 46: Cdt Free (Countdown Timer)

    - SET > go to Set CDT OFF, SET CDT SET MINUTES CDT SET SECONDS save save setting advance advance setting & return to minutes seconds CDT Status setpoint setpoint © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 47: Set Free Al (Alarms)

    Factory set for a fixed 30 seconds, the EDT alarm sounds the audible alarm every 30 seconds while underwater in Free Dive Mode. SET EDT AL ON or OFF toggle save setting setting © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 48: Depth Al (Alarm)

    > LIGHt INITIATING A DIVE With the i300 activated, a Free dive will commence upon descending to 5 ft (1.5 m) for at least 5 seconds. Below is a diagram to help you navigate Free Dive Mode functions. FREE DIVE MAIN...
  • Page 49: Free Dive Main

    FREE CDT (COUNTDOWN TIMER) ALARM When a set Countdown Time reaches 0:00, the audible alarm will sound. During which time, the graphic CDT will flash on the Free Dive Main screen. flashing © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 50: Free Edt (Elapsed Dive Time) Alarm

    VIO ALARM VIO ALARM N2 ALARM (during audible alarm) (after audible alarm) flashing flashing flashing flashing VIO < 1 MIN ON SURFACE VIO > 1 MIN ON SURFACE flashing alternating with FREE © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 51: Free Edt (Elapsed Dive Time) Alarm

    When set ON, the EDT alarm activates every 30 seconds during a dive. The audible alarm will sound. During which time, the graphic EDT and time digits will flash on the Free Dive Main screen. flashing flashing © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 52: Reference

    REFERENCE © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 53: Pc Interface

    PC INTERFACE The i300 is configured with a Data Port located on the side of the module that enables it to be connected to a PC through a USB port using a special interface cable that is available as an optional accessory. The USB Driver required for the interface system is downloadable from www.aqualung.com.
  • Page 54: Module Removal From A Boot

    • Send freight prepaid and insured using a traceable method. • Non-warranty service must be prepaid. COD is not accepted. • Additional information is available on the Aqua Lung web site AquaLung.com or on the local Aqua Lung web site that serves your global region.
  • Page 55 • Inspect the button, lens, and housing to ensure they are not cracked or damaged. WARNING: If damage or corrosion is found, return your i300 to an authorized Aqua Lung dealer, and DO NOT attempt to use it until it has received factory prescribed service.
  • Page 56: Returning The Module To A Boot

    • No adjustments are made during any time that the Wet Contacts are bridged. • When diving in high altitude waters from 3,001 to 14,000 feet (916 to 4,270 meters), the i300 automatically ad- justs to these conditions providing corrected depth, and reduced NO DECO and O2 MIN (O saturation) times at intervals of 1,000 feet (305 meters).
  • Page 57: Technical Data

    TECHNICAL DATA © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 58: No Deco Time Limits

    0:04 0:04 0:03 0:03 0:03 0:03 0:03 0:05 0:04 0:04 0:04 0:04 0:03 0:03 0:03 0:03 0:03 0:03 0:03 0:05 0:04 0:04 0:03 0:03 0:03 0:03 0:03 0:03 0:03 0:03 0:03 © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 59: Oxygen Exposure Limits

    OXYGEN EXPOSURE LIMITS (from NOAA Diving Manual) MAX DURATION MAX TOTAL DURATION SINGLE EXPOSURE 24 HOUR DAY (ATA) (MIN) (MIN) 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 60: Specifications

    • Out of the water - between 20 °F and 140 °F (-6.6 and 60 °C). • In the water - between 28 °F and 95 °F (-2.2 and 35 °C). © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 61 59:59 to 0:00 min:sec 1 second • Violation Countdown Timer 23:50 to 0:00 hr:min 1 minute Max Functional Depth: Limit: • DIVE/FREE 330 FT (100 M) • GAUGE 399 FT (120 M) © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 62: Abbreviations/Terms

    LAST = Previous (dive) LO = Low (battery) M = Meters (depth) MET = Metric MFD = Maximum Functional Depth (equipment limits) MIN = Minutes (time) MOD = Maximum Operating Depth © Aqua Lung International, Inc. (2015) Doc. 12-7823-r06 (12/10/15)
  • Page 63 www.aqualung.com...

Table of Contents