CAUTION
approved processes provided in this manual. Cleaning and disinfection methods listed are
recommended by Verathon based on efficacy or compatibility with component materials.
CAUTION
a power adapter or charging cradle. If the micro‑USB port is not dry when a power
adapter or charging cradle is connected to it, electric shock, equipment damage, or system
malfunctions can occur.
CAUTION
tools when cleaning or disinfecting the video monitor screen. The screen can be scratched,
permanently damaging the device.
CAUTION
Submerging the charging cradle in a liquid solution can cause system malfunctions or
damage to the monitor or the power adapter.
SYMBOLS
For a full list of caution, warning, and informational symbols used on this and other Verathon
products, please refer to the Verathon Symbol Directory at verathon.com/symbols.
This product may only be cleaned or disinfected by using the
Ensure the monitor's micro‑USB port is dry before connecting
Ensure that you do not use any abrasive brushes, pads, or
Do not submerge the charging cradle in a liquid solution.