[STATe:] ERRor
Command Syntax:
[ STATe:] ERRor ?{;NL}
Purpose: Query if there are any errors flagged in the module.
Description:
1. ERR? : Read the register of ERR status. TABLE 4-10 shows the
Corresponding number of ERR status
2. Use command CLR to clear the register of ERR status to be"0"
BIT ID
BIT VALUE
bit 0~3
0 = Off, 1 = Triggered
bit 4
0 = Off, 1 = Triggered
bit 5
0 = Off, 1 = Triggered
[STATe:] NG?
Command Syntax:
[ STATe:] NG? {;NL}
Purpose: Query if the NG flag is displayed on this module
Description: Set command NG? to show the NG status. If the response is "0" the LED
of NG(NO GOOD) will be off. If the response is "1" the LED will be lit, showing
that the NG flag is present.
[STATe:] PROTect?
Command Syntax:
[ STATe:] PROTect?{;NL}
Purpose: Query the state of the protection register on this module
Description: 1.PROT? requests the status of the units protection register.
2. Use the command "CLR" to clear the register of PROT status to "0"
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit1 Bit 0
7
6
5
4

TABLE 4-10 ERROR register

Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit1 Bit 0
7
6
5
4
3302F Series Mainframe Operation Manual 57
3
2
1
0
ERROR OPERATION
ERROR COMMAND
REMARK
CH1 error
Operation error
Command error (e.g. syntax error)
3
2
1
0
Over Voltage Protection (OVP)
Over Current Protection (OCP)
CH1 ERROR
Over Power Protection (OPP)
Over Temperature Protection (OTP)