Appendix 1 - Battery Replacement
Prepare the Recorder
Download the unit to save any recorded data.
l
Hibernate the unit using LogTag
l
Failing to hibernate the recorder may result in serious memory
corruption. As a result you may no longer be able to use this
product.
Remove the old battery
Carefully remove the rear label covering the battery door using a
l
small-bladed screwdriver or knife:
Use a coin to turn the battery door anti-clockwise; remove it from
l
the case:
Remove the battery using a small-bladed screwdriver, as shown:
l
Distributed by MicroDAQ
Page 17 of 17
®
Analyzer.
www.MicroDAQ.com
Insert the new battery
Only use batteries from a reputable manufacturer! Check the
operational temperature range of the battery and make sure it
covers the range of the TRED30-16R.
Put a new CR2032 cell into the battery compartment as shown;
l
the right hand edge of the battery is inserted first against the
contact:
Press firmly down on the left hand side of the battery to click the
l
battery into place:
Replace the battery door; turn clockwise to lock it into place.
l
Cover the battery door with either a new label (Part # 100-
l
000502) or reuse the old one.
Re-configure the recorder with LogTag
l
Please recycle or dispose of the old battery according to your local
environmental regulations.