Page 3
… the device is stored and operated according to LogTag Recorders ’ recommendations. Disclaimer The TRED30-16R monitors temperature exposure and not the quality of the goods it accompanies. Its purpose is to signal if product quality evaluation/testing is required. Typographical Conventions Text refers to buttons on the TRED30-16R.
Page 4
The TRED30-16R temperature recorder features the familiar LogTag case layout. Configuring the TRED30-16R Before a TRED30-16R recorder can be used, it must be configured with the parameters required for starting and recording temperature ® values. This is done using LogTag Analyzer software, which is also used for downloading and analyzing data.
Page 5
( ) to If you configure the TRED30-16R to start after a delay period, the indicate the alarm was registered against what is now the previous recorder will not immediately record temperature readings after day.
Page 6
Audible Alarm Advanced Configuration Options The TRED30-16R is fitted with a beeper, which can be turned on Select Advanced Options for additional configuration settings. These when an alarm event has been registered to provide extra feedback. settings decide how some of the elements are displayed on the screen This function is enabled or disabled when configuring the recorder and set certain options specific to the TRED30-16R.
Page 7
The word is shown in Review mode, when the temperature on symbol is shown as soon as the TRED30-16R has registered an the display represents the maximum recorded temperature for the day alarm event. If no alarms have been registered, or if an existing alarm displayed.
Page 10
Automatic date/time start If the recorder has been configured for push button start with a start If you configured the TRED30-16R for a date/time start, it will start recording temperature values as soon as the entered start time is delay (regardless of quickstart settings), the word...
Page 11
REVIEW /MARK During normal operation the display shows the most recently recorded by the TRED30-16R will be identified in the downloaded data with an temperature. This temperature is updated at the same rate as the inspection mark. logging interval. The current time is also displayed (in 24 hour format).
Page 12
This function is useful if the sensor is permanently Figure 6: TRED30-16R recording, showing ‘18’ days of data has been collected. mounted , but you wish to take the TRED30-16R to a computer for downloading its data. Figure 7: TRED30-16R stopped, showing ‘27’ days of data has been collected.
Page 14
-07 is shown to indicate that the data DAYS You can configure a TRED30-16R so it can be stopped with the being displayed is from seven days ago. Seven days ago recordings above the upper threshold were recorded, button.
Page 16
All ST100 precision thermistor external probes can be interchangeably connected Battery life ® to the TRED30-16R. Please refer to the dedicated sensor page on the LogTag Typically 1 year of operation with normal use (6 minute logging, statistics Recorders website.
Need help?
Do you have a question about the TRED30-16R and is the answer not in the manual?
Questions and answers