Vega VEGABAR 87 Operating Instructions Manual page 53

Submersible pressure transmitter with metal measuring cell
Hide thumbs Also See for VEGABAR 87:
Table of Contents

Advertisement

Verbindungskabel bei IP68 (25 bar)-Ausführung
Ʋ Kabelmantel
Ʋ Typschildträger auf Kabel
Werkstoffe Messwertaufnehmerschutz
Transport- und Montageschutz
Transportschutznetz
Gewichte
Grundgewicht
Tragkabel
Verbindungsrohr
Abspannklemme
Verschraubung
Torques
Max. torque for process fitting
Ʋ G1½
Max. torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing
Ʋ Aluminium/Stainless steel housing
Input variable
The specifications are only an overview and refer to the measuring cell. Limitations due to the
material and version of the process fitting as well as the selected pressure type are possible. The
specifications on the nameplate apply.
Nominal measuring ranges and overload capability in bar/kPa
Nominal range
Gauge pressure
0 ... +0.1 bar/0 ... +10 kPa
0 ... +0.4 bar/0 ... +40 kPa
0 ... +1 bar/0 ... +100 kPa
0 ... +2.5 bar/0 ... +250 kPa
0 ... +10 bar/0 ... +1000 kPa
0 ... +25 bar/0 ... +2500 kPa
Absolute pressure
0 ... 1 bar/0 ... 100 kPa
0 ... 2.5 bar/0 ... 250 kPa
0 ... 10 bar/0 ... 1000 kPa
Zwischen Messwertaufnehmer und externem Elektronikgehäuse.
4)
Data on overload capability apply for reference temperature.
5)
VEGABAR 87 • 4 ... 20 mA
4)
PE, PUR
PE-hart
PFA
PE
0,7 kg (1.543 lbs)
0,1 kg/m (0.07 lbs/ft)
1,5 kg/m (1 lbs/ft)
0,2 kg (0.441 lbs)
0,4 kg (0.882 lbs)
200 Nm (147.5 lbf ft)
10 Nm (7.376 lbf ft)
50 Nm (36.88 lbf ft)
5)
Maximum pressure
+15 bar/+1500 kPa
+25 bar/+2500 kPa
+25 bar/+2500 kPa
+25 bar/+2500 kPa
+25 bar/+2500 kPa
+25 bar/+2500 kPa
25 bar/+2500 kPa
25 bar/+2500 kPa
25 bar/+2500 kPa
11 Supplement
Overload capability
Minimum pressure
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
0 bar abs.
0 bar abs.
0 bar abs.
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vegabar 86

Table of Contents