6 Set up with the display and adjustment module
Menu item
Adjustment
Damping
Linearisation
Current output
Lock adjustment
Display
Menu item
Menu language
Displayed value 1
Displayed value 2
Display format 1 and 2
Backlight
Diagnostics
Menu item
Sensor status
Peak value indicator
Simulation
Additional adjustments
Menu item
PIN
Date/Time
Copy instrument settings
Special parameters
38
Parameter
Zero/Min. adjustment
Span/Max. adjustment
Integration time
Current output - Mode
Current output - Min./Max. 3.8 mA
Default value
Selected language
Current output in %
Ceramic measuring cell: Measuring cell temperature in °C
Metallic measuring cell: Electronics temperature in °C
Number of positions after the decimal point, automatically
Switched on
Parameter
Pressure
Temperature
Parameter
Default value
0.00 bar
0.00 %
Nominal measuring range in bar
100.00 %
1 s
Linear
Output characteristics
4 ... 20 mA
Reaction when malfunctions occur
≤ 3.6 mA
20.5 mA
Released
Default value
-
Actual measured value
Actual temperature values from measuring cell, elec-
tronics
Process pressure
Default value
0000
Actual date/Actual time
No reset
VEGABAR 87 • 4 ... 20 mA