Event
Description
ID
to insufficient optical signal
222
Q-Factor out of range
Q factor < 0.5 or Q factor > 1.50
Failed reading dark gray
223
Optics Verification Failed
leader
Failed reading light gray
224
Optics Verification Failed
leader
Optics blocks fail at SPI
No communication with Optics for more than 5
225
communication
seconds
Optics reference
Reference photodiode output exceed operating
226
photodiode out of range
range during Optics adjustment
LED Drive not stable in
Optics block reports unstable reading 10 times.
227
Optics
Close-loop control for LED drive not stable
Control module not
Analyzer unable to communicate with the Single
228
responding to Analyzer
Board Computer for 10 seconds
PDU not responding to
Analyzer unable to communicate with the Power
229
Analyzer
Distribution Unit (PDU) for 10 seconds
No valid Chemcassette
No valid RFID tag detected when the Analyzers
230
detected
starts monitoring
Failed to write data on the tag 5times
231
Chemcassette write failure
consecutively
Set Condition
Optics LED not properly calibrated
LED degradated
Optics board defective
Chemcassette leader not tight or improperly
positioned during white to light gray calibration
Bad RFID tag
Dirty optics block
Bad optics PCB set
Chemcassette leader not tight or improperly
positioned during white to light gray calibration
Bad RFID tag
Dirty optics block
Bad optics PCB
Chemcassette leader not tight or improperly
positioned during light gray to dark gray
calibration
Bad RFID tag
Dirty optics block
Bad optics PCB
Bad optics PCB
Dirty Optics block
Hardware failure
Hardware failure
Hardware failure
Hardware failure
Chemcassette changed without using Change
Tape utility
Non-Honeywell tape installed
RFID Reader
Possible Cause
Contact Honeywell Analytics Service.
Perform Load CC Operation to recalibrate
Replace the optics block
Replace Analyzer, Contact Honeywell service
Reload Chemcassette
Load new Chemcassette
Clean and recalibrate
Service or replace analyzer
Reload Chemcassette
Load new Chemcassette
Clean and recalibrate
Service or replace analyzer
Reload Chemcassette
Load new Chemcassette
Clean and recalibrate
Service or replace analyzer
Replace Optics Block
Service or replace analyzer
Clean optics.
Contact Honeywell Analytics Service.
Contact Honeywell Analytics Service
Contact Honeywell Analytics Service
Contact Honeywell Analytics Service
Contact Honeywell Analytics Service
Reload Chemcassette
Unauthorized Tape installed
Contact Honeywell Analytics Service
Replace RIFD Board
Replace Analyzer, Contact Honeywell service
Resolution