Honeywell Vertex Edge User Manual page 92

72-point continuous monitor
Hide thumbs Also See for Vertex Edge:
Table of Contents

Advertisement

Event
Description
ID
Gas table file is bad or
209
Gas table not programmed or CRC error
missing
210
Optics drive unusually low
Optics LED drive < 1000
211
Optics drive unusually high
Optics LED drive > 4000
212
Excess optics signal
Optics Primary Counts > 3700
213
Double Pump failure
Double Pump Failure
Point flow is below 20cc/min (loss of flow) in
214
Loss of Flow
monitoring mode.
Failed to stop Analyzers
215
Analyzers won't stop monitoring mode
monitoring mode
Point Pressure Out Of
Point pressure excceed 30" Hg for more than 5
216
Range
seconds during monitoring mode
System Pressure Out Of
System pressure exceed 30"Hg for more than 5
217
Range
seconds during monitoring mode
System pressure is lower than 6.0"Hg for more than
Inadequate Analyzer
60 seconds when there is no pump failure.
218
Vacuum
Individual analyzer observes inadequate vacuum,
stops monitoring.
219
Optics SW corrupted
Optics block software not communicating
220
Optics Internal SW errors
Optics block reports incorrect information
221
LED adjustment failed due
LED adjustment failed (reported by Optics)
Set Condition
No configuration loaded
Optics LED not properly calibrated
Optics board defective
Analyzer CPU defective
Optics LED not properly calibrated
LED degradated
Optics board defective
Analyzer CPU defective
Optics LED not properly calibrated
Optics board defective
Analyzer CPU defective
See causes for Fault 109
Circuit breaker tripped
Improper flow setup
Tape tracking problem
Condensation in system
Clogged micro tube
Gate not fully closing
Optics block loose
Flow adjustment is unstable during adjusting
Queen Analyzer (first in the rack) failed to
communicate with other Analyzers
Analyzer hardware failure
Miscalibrated sensor board or defective
transducer
Miscalibrated sensor board or defective
transducer
Improper system pressure adjustment
Hardware failure
Cable issue
Hardware failure
Optic block dirty
Possible Cause
Reinstall Configuration Profile
Perform Load CC Operation to recalibrate
Replace Analyzer, Contact Honeywell service
Replace Analyzer, Contact Honeywell service
Perform Load CC Operation to recalibrate
Replace the optics block
Replace Analyzer, Contact Honeywell service
Replace Analyzer, Contact Honeywell service
Perform Load CC Operation to recalibrate
Replace Analyzer, Contact Honeywell service
Replace Analyzer, Contact Honeywell service
See recoveries for Fault 109
Contact Honeywell
Adjust flow to 200cc
Reload Chemcassette
Purge internal lines
Replace tube.
Note: Calibration required, contact Service.
Gate adjustment loose, pivot binding
Tighten optics block fasteners
Replace needle valves
Replace first Analyzer in the rack
Replace Analyzer, Contact Honeywell service
Replace Analyzer
Replace Analyzer
Adjust system pressure to 10"Hg
Replace Analyzer
Confirm cable connections
Replace Analyzer, Contact Honeywell service
Clean optics.
Resolution