Honeywell Vertex Edge User Manual page 90

72-point continuous monitor
Hide thumbs Also See for Vertex Edge:
Table of Contents

Advertisement

Event
Description
ID
Expired
Optics Cleaning
119
Optics cleaning time in configuration profile elapsed
Timer Expired
Pump
120
Maintenance
Pump rebuild time in configuration profile elapsed
Timer Expired
*Reserved for LIT
121
option
*Reserved for LIT
122
option
*Reserved for LIT
123
option
*Reserved for LIT
124
option
Possible debris in
125
Optics signal has been erratic on at least 3 windows
optics block
Abnormal Optics
Compensated abnormal optics reading and keep recode for information
126
reading detected
when the option of "Generate window zero reset" is enabled
Accelerated
127
Chemcassette
Daily tape usage is higher than 2 times estimated tape usage per day
Usage
Ethernet
128
Analyzer ethernet initialization failed during boot up
initialization failed
File system of
129
Analyzer is
NorFlash - File system error
corrupted
Optics blocks have
130
different software
The two optics blocks are not running the same software version.
versions
Unable To Log
131
Analyzer event log function failed
event data
Software version
132
mismatched
In a rack, there is at least one analyzer with different SW version
among Analyzers
Set Condition
Possible Cause
Maintenance reminder, no malfunction
Maintenance reminder, no malfunction
Debris in optics block
Compensated optic reading automatically. So no
further action required. But too frequently happens,
debris in the optic block.
Low level background gas below lower detectable
limit
Failed load the driver, Electronic problem
File system corrupted
Program update done incorrectly
Electronic problem
New analyzer was installed into the Vertex Edge rack
that contains a different software revision than the
other analyzers
Resolution
Clean the Optics and reset timer
Rebuild pump and reset timer
Clean optics block
Clean optics block
Locate source of background gas
Reboot the analyzers.
Contact Honeywell Analytics Service.
Contact Honeywell Analytics Service
Reload program to both optics blocks.
Contact Honeywell Analytics Service
Upgrade analyzer Software