7. OPERATION
7.1. START UP OF MANIFOLD
When the unit is completely installed
and all shut off valves are closed.
Slowly open gas cylinder shut-off
valves.
Slowly open shut-off valves V1, V2
and V3.
Open purge valve V5.
Slowly open main-shut-off valve V4
to ventilate the unwanted "air" from
EN
the collecting unit and high pressure
hoses and close it back.
Close purge valve V5. The purge valve V5 must be closed during operation.
You should keep at least one of the cylinders closed to be used as reserve one. Close shut-off
valve V1, V2 and/or V3 for concerned cylinder(s) and concerned gas cylinder shut-off valve.
Assign cylinder with indication panels (Fig. 4).
To check if there is enough gas in the cylinders, use inlet gauge M1.
Check gauge M2 if the outlet pressure is as required.
7.2. CYLINDER CHANGE OVER
When 3 gas cylinders are connected to the collecting unit C1 and the one which is in use
indicates low pressure,the following steps are to be taken:
SLOWLY open full cylinder and its corresponding shut off valve on the collecting unit.
Close empty cylinder and the corresponding shut-off valve.
The gas cylinder indication panels must be now adjusted so that it is clearly apparent which
cylinder is in use.
Exchange of the empty cylinder(s) for a full one.
Close shut-off valves (V1 and/or V2 and/or V3 and/or corresponding to full gas cylinders or
cylinders you do not wish to exchange. The gas flow supply has now been stopped.
Open shut-off valve V4.
Close cylinder shut off valve of the empty cylinder(s) that you wish to exchange.
Open purge valve V5 to evacuate remaining gas from the collecting unit and high pressure
hoses. Close the purge valve V5 when all gas is evacuated.
Close the opened shut-off valves (V1 and/or V2 and/or V3 and V4).
Exchange empty cylinder(s) with full one.
Ventilate the unwanted gas as described in chapter 7.1 from the changed cylinders.
Then continue using the gas supply from the previously used cylinders. The indication panels
must always match the cylinder(s) in use to prevent unexpected running out of gas.
When 3 gas cylinder are connected to the collecting unit C1 and all 3 are empty, the following
steps are to be taken:
Open main shut off valve V4 .
Close cylinder shut off valves.
Open purge valve V5 to evacuate remaining gas from the collecting unit and high pressure
hoses. Close the purge valve V5 when all gas is evacuated.
Close the shut off valves V1, V2, V3 and V4.
Exchange empty cylinders for full ones.
Now follow steps from chapter 7.1 - start up the manifold.
7.3. SHUT DOWN
The manifold can be only shut down with agreement of a responsible person according to
standard EN ISO 7396-1.
For short term (hours) shutdown close shut-off valve V4.
For long term (days) shutdown close all opened shut-off valves on the unit (V1 - V4)and the
cylinder shut-off valves
Indication of gas type
Indication panels
Fig. 4: Indication panels