Shower / disinfecting unit
Table of Contents

Advertisement

Adjusting the flow meter
1. Make sure that the hose, that is also connected to the flow meter, is fully immersed in
the bottle with disinfectant. Otherwise, no disinfectant will be sucked through and the
flow meter cannot be adjusted.
2. On the packaging of the disinfectant, look for the recommended mixture ratio. Usually
this is shown in millilitres per millilitres (for example: 10ml / 1000ml). If not, this will
need to be converted.
3. Now establish the time in which it has to be measured how much water is running out
of the tap.
You do this by dividing 3600 by the recommended mixture ratio.
EXAMPLE: If there is a mixture ratio of 10ml / 1000ml, 100 parts of water are mixed to
1 part of disinfectant. The mixture ratio in this case is 100. When you divide 3600 by 100,
you get 36. This figure is the number of required seconds for step 3.
4. Open the disinfectant tap and safety tap. Now press the black button on the red hand-
held shower in fully for the exact number of seconds calculated in step 3. Measure the
number of litres that have flowed from the disinfecting shower during that time.
5. This amount in litres is the same value that the flow meter has to show when the
black button of the red hand-held shower is fully pressed and water is running through
it. When the black button on the flow meter is turned, the correct level can be set.
Please note: if you are NOT certain that you understand the aforementioned
method or if you feel you have not applied it correctly, please ask your
manager for assistance or phone Lopital Nederland B.V.