Replace or refill disinfectant
• It is advised to wear plastic gloves when adding or replacing the disinfectant.
• The storage bottle for the disinfectant is stored behind the locked flap.
• Open the flap; turn the key a quarter turn to the left and tip the flap forwards.
• Turn the lid of the bottle, watch out for the hose!
• Insert the hose into the new bottle and place the old lid on the new bottle
• Close the flap.
• Lock the flap by turning the key one quarter turn to the right.
• Rinse out the empty bottle properly and dispose of with the refuse. Bottles which still
contain a substance have to be disposed of with the chemical waste.
• If a different disinfectant is used, the flow meter has to be readjusted.
• Frequently change the disinfectant liquid to keep the disinfecting capacity optimal.
Lopital recommends the use of Neoquat-comb
as a disinfectant.
PLEASE NOTE: if a different disinfectant is used to replenish the reservoir,
the LD50 level of that liquid may not be lower than 200 mg / kg! If this is
lower, the appliance will be unsafe and use of the appliance is totally at the
user's own risk!
Lopital recommends that the entire bottle is replaced and not replenished, in order to
avoid mixing different disinfectants together. Mixing could possibly cause undesirable
reactions between the liquids and between the liquids and the appliance.