  • Page 1 Instructions for use Shower / disinfecting unit ‘Amfora’...
  • Page 2 English Important: Read these instructions carefully before using the Amfora! Date of first use User Name Address Manufactured by: Lopital Nederland B.V. Laarakkerweg 9, 5061 JR Nederland T (0031) 13 52 39 300 F (0031) 13 52 39 301 www.lopital.nl...
  • Page 3 Replace of refill disinfectant Replace or refill shampoo and soap Maintenance Technical information Soap pump Shower tap Disinfectant shower (red) Hand-held shower Disinfectant tap (white) Flow meter Thermostat Disinfectant Flap Safety tap Shower / disinfecting unit ‘Amfora’ | Instructions for use | 3...
  • Page 4 Introduction The Amfora shower and disinfecting unit is a combined appliance for washing people and disinfecting, amongst others, shower trolleys and shower and toilet chairs. The Amfora is a product with CE marking. This marking is in accordance with guideline 93 / 42 / EEC, and respectively the revised guideline 2007 / 47 / EC for medical appliances of March 2010.
  • Page 5: Adjusting The Temperature

    • If the correct taps are opened on the Amfora, pressing the black button will cause the water to flow from the shower head. By pressing the button further in, or by pressing it less the amount of water can be regulated.
  • Page 6 • Allow the disinfectant to soak in for the specified amount of time (for Neoquat-Combi that is 5 minutes). ® • Open the shower tap and remove the white hand-held shower from its holder. 6 | Instructions for use | Shower / disinfecting unit ‘Amfora’...
  • Page 7 Dose too low Low disinfecting effect, build up resistance Dose too high Residue problems, corrosion Disinfectant not suitable Inadequate disinfection, discolouration Moisture present before disinfecting Change of concentration of disinfectant Shower / disinfecting unit ‘Amfora’ | Instructions for use | 7...
  • Page 8: Flow Meter

    The flow meter has to be adjusted when the Amfora is being installed, and each time that you are intending to use a different type or brand of disinfectant. It will be easier to reach the flow meter if you remove the left-hand reservoir of the soap pump.
  • Page 9 Please note: if you are NOT certain that you understand the aforementioned method or if you feel you have not applied it correctly, please ask your manager for assistance or phone Lopital Nederland B.V. Shower / disinfecting unit ‘Amfora’ | Instructions for use | 9...
  • Page 10 Lopital recommends that the entire bottle is replaced and not replenished, in order to avoid mixing different disinfectants together. Mixing could possibly cause undesirable reactions between the liquids and between the liquids and the appliance.
  • Page 11: Maintenance

    • Lopital Nederland B.V. recommends that an annual maintenance contract is entered into for the Amfora. • Maintenance and repair should only be performed by authorised personnel. The Amfora has to be inspected and maintained at least once a year. Shower / disinfecting unit ‘Amfora’ | Instructions for use | 11...
  • Page 12: Technical Information

    Dimensions / weights External dimensions 55 x 63 x17 cm (w x h x d) (excl. hoses) Empty weight 12 kg Volume of disinfectant Volume of shampoo 2x 0.5 l 12 | Instructions for use | Shower / disinfecting unit ‘Amfora’...
  • Page 13 For more information, please visit our website: www.lopital.com Lopital Nederland B.V. Laarakkerweg 9, 5061 JR Oisterwijk, Postbus 56, 5060 AB Oisterwijk Tel 013 5239300, Fax 013 5239301, E-mail info@lopital.nl, www.lopital.nl For more information, please visit our website: www.lopital.com Lopital België Bvba, Antwerpsesteenweg 124, 2630 Aartselaar...