12.2 Manual Cell Clean Unit (MCU)

Optional item if fitted
The Manual Cleaning Unit (MCU) facilitates cleaning of the cell without the need
of removing the cell cap. Regular use of this device should prevent malfunction of
the monitor due to common fouling mechanisms of the sample tube, and may
allow longer maintenance intervals. The MCU (Order No. 77780) can be installed
at any time.
Operating Instructions
a) Ensure that there is a clean water supply through the cell.
b) Activate the manual cell clean unit by pressing the handle several times.
c) Check the reading is between 0 to 2 ppm.
d) Repeat a) to c) at least once a week or as necessary.
e) Additional conventional cleaning with the cleaning brush may be
necessary.
Issue: 25.09.2015 preliminaryInstruction Manual OMD-2008 (All Versions)