SIGNALCORE SC5511A Operating & Programming Manual page 48

100 mhz to 20 ghz rf signal source
Table of Contents

Advertisement

C
& M
A L I B R A T I O N
A I N T E N A N C E
The SC5511A is factory calibrated and ships with a certificate of calibration. SignalCore strongly
recommends that the SC5511A be returned for factory calibration every 12 months or whenever a
problem is suspected. The specific calibration interval is left to the end user and is dependent upon the
accuracy required for a particular application.
Should any customer need to reload calibration data for their SC5511A, SignalCore offers free support
through support@signalcore.com. SignalCore will provide a copy of the archived calibration data along
with instructions on how to upload the file to the SC5511A.
The SC5511A requires no scheduled preventative maintenance other than maintaining clean, reliable
connections to the device as mentioned in the Getting Started section of this manual. There are no
serviceable parts or hardware adjustments that can be made by the user.
SC5511A User Manual
Rev 1.7.0
44